Keysight 53210A/53220A/53230A Assembly Level Service Guide
Device-specific errors (-300..., +800)
These error messages can be either operator information only, or can indicate a
possible hardware failure.
– -310, “System error; internal software error”
– (1) P0500-63001 Processor board
– -310, “System error; software initialization failed”
– (1) P0500-63001 Processor board
– -310, “System error; out of memory”
– (1) P0500-63001 Processor board
– -311, “Internal software error”
– -313, “Calibration memory lost”
– (1) P0500-63001 Processor board
– -313, “Calibration memory lost; memory corruption detected”
– (1) P0500-63001 Processor board
– -313, “Calibration memory lost; due to firmware revision change”
– -314, “Save/recall memory lost; memory corruption detected”
– (1) P0500-63001 Processor board
– -314, “Save/recall memory lost; due to firmware revision change”
– -315, “Configuration memory lost; memory corruption detected”
– (1) P0500-63001 Processor board
– -315, “Configuration memory lost; due to firmware revision change”
– -330, “Self-test failed”
– -350, “Error queue overflow”
– +800, “Nonvolatile memory write failure”
– P0500-63001 processor board
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...