2
Service
144
Keysight 53210A/53220A/53230A Assembly Level Service Guide
– After a few seconds, display momentarily shows:
15
Select the ‘output enable’ ON feature of the microwave source or reconnect
the cable to the counter, as appropriate.
16
Use the front panel knob and left/right arrow keys to enter the value (300 MHz)
as shown.
17
Press the
Input Power
softkey and enter the power level of the
300 MHz source as -21 dBm. (Actual input power at the counter must be -21
dBm ± 0.5 dB. If necessary, verify with a power meter.)
18
Press the
Perform Cal
Step
softkey to enter the data into the counter and initiate
the microwave oscillator
calibration.
– If the calibration step PASSES, after a few seconds the counter momentarily
displays “Calibration step succeeded.”
– The internal calibration constants are updated.
– The
Cal Count
on the screen will increment.
19
Repeat steps 16 to 18 using each of the calibration frequencies and power
(total of 36 steps). For each calibration step, set the source
frequency and power accordingly.