2
Service
118
Keysight 53210A/53220A/53230A Assembly Level Service Guide
The Keysight 53210A/53220A/53230A Calibration Procedures
Equipment required
Table 2-1
Recommended test equipment for calibration
Instrument
Required characteristics
Recommended model
10 MHz system clock or
RF signal generator
10 MHz ± 2.000E-12
Stanford Research Systems CG635
or equivalent
Primary frequency standard
(10 MHz house standard)
Absolute accuracy >1 x 10-10
Stanford Research Systems FS725
Rubidium Timebase or equivalent
Digital multimeter (DMM)
Microvolt accuracy
Keysight 34410A or equivalent
Microwave signal generator
100 MHz to 15 GHz; Freq Accy at 6
GHz and 15 GHz:
±10ppb;
PW Accy:
±10%
Keysight E8257D-520-UNW or
equivalent
DC power supply
Adjustable to ± 5.000V
Keysight 6234A or equivalent
Type N(m)-to-BNC(f) Adapter
Keysight 1250-0780
SMA(m)-to-BNC(f) adapter
Keysight 1250-1200
BNC-to-dual banana plug
BNC(f) to banana plug(m)
Keysight E9637A or Newark
10M4161
50 Ohm coaxial cable with
BNC connectors (5)
BNC(m) to BNC(m), 48 inches
Keysight 10503A
3.5mm cables
DC-18 GHz, matched pair
Keysight 11500E
Coaxial BNC tee connector
Keysight 1250-0781
Some of this equipment may no longer be available as new.
Use an equivalent model number.
NOTE
Whenever possible, the counter should be calibrated with the instrument
oriented in the same physical direction in which it will be used (i.e. flat, on one
side, or front panel facing up).
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...