1
Performance Tests
70
Keysight 53210A/53220A/53230A Assembly Level Service Guide
7
Press the
Trigger
key.
– The counter takes 100 readings and fills in the math functions.
8
Verify the ‘Mean’ is:
– 5.88 GHz ± 58.800 Hz (53210A)
– 5.88 GHz ± 5.880 Hz (53220A/53230A)
9
Verify the ‘Std Dev’ is less than:
– 82.300 Hz (53210A)
– 41.100 Hz (53220A)
– 8.230 Hz (53230A)
10
Record the Mean and Standard Deviation for 5.88 GHz, Test 6a,
in the Keysight 53210A/53220A/53230A Performance Test Record.
11
Press the
Reset Stats
softkey.
12
If Opt 150 is installed, proceed to Test 7a. Otherwise, disconnect the test
setup.