1
Performance Tests
58
Keysight 53210A/53220A/53230A Assembly Level Service Guide
8
Press the Channel
1
key:
– The Channel
1
Setup softkeys are displayed:
9
Press and set:
– Coupling:
DC
– Input Impedance:
50
Ω
– Range:
5 V
– BW Limit:
OFF
10
Press the
Math
key.
– The following softkeys are displayed:
11
Press and set:
– All math:
On
12
Press the
Statistics
softkey:
– The
Statistics
softkeys are displayed: