Service
2
Keysight 53210A/53220A/53230A Assembly Level Service Guide
129
–
Cal String
shows the current calibration string. This string is
operator-modifiable, used to record such information as the date of the last
calibration, or by whom it was performed. The Calibration String can not be
entered from the front panel keys. It can only be entered/updated using the
remote SCPI command:
CALibration STRing
.
Refer to the Keysight
53210A/53220A/53230A SCPI Programmer’s Reference for details of the
command format.
–
Secure Code
displays either all dashes ( --- ) or an operator-entered security
code that has not yet been accepted by the counter.
5
When
“Lock Cal”
is displayed for a softkey, the firmware is unlocked and ready
for calibration. The
Perform Cal
softkey then becomes available.
6
Press the
Perform Cal
softkey.
– Display softkey annunciators show:
7
Press the
Cal Step
softkey.
– Display softkey annunciators show:
* Only displayed if Option 106 or 115 is installed.
NOTE
If
“Unlock Cal”
is displayed for a softkey, the firmware is locked against
calibration. (See step 2.)
*