Performance Tests
1
Keysight 53210A/53220A/53230A Assembly Level Service Guide
45
5
Press the
Freq
softkey.
– The following softkeys are displayed:
6
Press and set:
– Noise Rejection:
Off
7
Press the
Trigger
key.
– The following softkeys are displayed:
8
Press and set:
– Trigger Source:
Manual
–
Wait For Trig
is displayed.
– Rdgs/Trigger:
100
9
Press the
Gate
key.
– The
Gate
softkeys are displayed:
10
Press and set:
– Gate src:
Timed
– Gate time:
10 mSec
– Gate out:
On
– Polarity:
Norm