Press the
Input Power
softkey and enter the power level of the 500 MHz source
as -27 dBm (actual input power at the counter must be -27 dBm ± 0.5 dB).
16
Press the
Perform Cal
Step
softkey to enter the data into the counter and initiate
the microwave oscillator
calibration.
– If the calibration step PASSES, after a few seconds the counter momentarily
displays “Calibration step succeeded.”
– The internal calibration constants are updated.
– The
Cal Count
on the screen will increment.
17
Repeat steps 14 to 16 using each of the calibration frequencies and power
levels in
(total of 36 steps). For each calibration step, set the source
frequency and power accordingly.
Table 2-2
Option 106 calibration frequencies and power levels
Test point
Frequency (MHz)
Power level 1
Power level 2
Power level 3
1
500 MHz
-27 dBm
-4 dBm
+10 dBm
2
1500 MHz
-27 dBm
-4 dBm
+10 dBm
3
2500 MHz
-27 dBm
-4 dBm
+10 dBm
4
3000 MHz
-27 dBm
-4 dBm
+10 dBm
5
3500 MHz
-27 dBm
-4 dBm
+10 dBm
6
4000 MHz
-27 dBm
-4 dBm
+10 dBm
7
4500 MHz
-27 dBm
-4 dBm
+10 dBm
8
5000 MHz
-27 dBm
-4 dBm
+10 dBm
9
5250 MHz
-27 dBm
-4 dBm
+10 dBm
10
5500 MHz
-27 dBm
-4 dBm
+10 dBm
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...