Keysight 53210A/53220A/53230A Assembly Level Service Guide
18
If any calibration step FAILS, after several seconds the counter displays
“Calibration step failed”
.
19
If the calibration fails (“
Calibration step failed
”), first verify the input frequency
and power are correct and the source output is connected and turned on.
Then, refer to
20
To check for error messages, refer to the
below for information on how to read the error
message(s) from a remote computer.
21
If you are able to read the error message(s), consult
page 174 for what to do after receiving a specific calibration error message.
11
5750 MHz
-27 dBm
-4 dBm
+10 dBm
12
6000 MHz
-27 dBm
-4 dBm
+10 dBm
Table 2-2
Option 106 calibration frequencies and power levels (continued)
Test point
Frequency (MHz)
Power level 1
Power level 2
Power level 3
NOTE
If the calibration fails for any calibration point (including zero-cal), the current
calibration constants for that frequency and power level, and the
Cal Count,
remain unchanged.
NOTE
If you do not have a remote computer available (requires a LAN, USB or GPIB
interface) or are not proficient at programming, contact the nearest Keysight
Service Center for information and help on sending the counter to Keysight for
repair and/or calibration.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...