Service
2
Keysight 53210A/53220A/53230A Assembly Level Service Guide
127
Voltage calibration (internal DACs)
1
First, run the counter Self-Test as described under
on page 169. A successful self-test procedure verifies that the
counter is operational and ready for calibration.
2
Unlock the counter by performing the procedure
[Utility] / (Cal AutoCal) / enter <code> / (Unlock Cal )
3
Set up an external, low-noise +/- DC voltage source, such as the Keysight
6234A DC Power Supply, to output a voltage of +4.0 VDC. Do not connect the
output to the counter input connector until instructed to do so.
NOTE
An AutoCal is performed automatically after this calibration is successfully
completed.
NOTE
The Calibration menu is accessed by pressing the
Utility
key and then the
Cal/
AutoCal
softkey.
NOTE
The preferred voltage for optimum calibration accuracy is +4.096 Volts DC.
However, if your power supply can not be set to that specific value, set it as
close to +4.0 VDC as possible (acceptable range is 3-5 V). The entered value for
calibration MUST be within 1.0 milliVolt of the actual voltage source output. It is
suggested that a BNC ‘T’ be connected to the power supply output to enable a
DMM to measure the voltage that will be applied to the counter.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...