Performance Tests
1
Keysight 53210A/53220A/53230A Assembly Level Service Guide
39
Counter setup
1
Connect the equipment as shown in
Figure 1-3
Absolute timebase test setup
2
Press the
Preset
key on the front panel to preset the counter.
– Display shows:
–
Instrument has been preset.
– Channel 1 annunciator is lit.
3
Press the
Trigger
key.
– The following softkeys are displayed:
4
Press and set:
– Trigger Source:
Manual
–
Wait For Trig
is displayed.
– Rdgs/Trigger:
100.
FS725
Rubidium
Timebase
CG635
Signal
Generator
Ch 1