1
Performance Tests
54
Keysight 53210A/53220A/53230A Assembly Level Service Guide
8
Verify the ‘Mean’ is:
– 4.00 nSec ± 0.200 nSec (53220A)
– 4.00 nSec ± 0.100 nSec (53230A)
9
Verify the ‘Std Dev’ is less than:
– 0.140 nSec (53220A)
– 0.028 nSec (53230A)
10
Record the ‘Mean’ and ‘Std Dev’ for 250 MHz, Channel 1, Test 3b, in the
Keysight 53220A/53230A Performance Test Record starting on
11
Press the
Reset Stats
softkey.
12
Repeat steps 1 to 11 for Channel 2. Be sure to perform the Channel 2 setup
(see steps 6 and 7 under “Counter Setup”.)
13
Record the ‘Mean’ and ‘Std Dev’ for 1 MHz and 250 MHz, Channel 2, Test 3c
and 3d, in the Keysight 53220A/53230A Performance Test Record starting on
14
Press the
Reset Stats
softkey.
15
Proceed to Test #4 or disconnect the setup.
NOTE
Selecting Channel 2 automatically resets the statistics.