2
Service
136
Keysight 53210A/53220A/53230A Assembly Level Service Guide
12
Press the
Perform Zero-Cal
softkey to perform zero-calibration.
– Display momentarily shows:
– After a few seconds, display momentarily shows:
13
Select the ‘output enable’ ON feature of the microwave source or reconnect
the cable to the counter, as appropriate.
14
Use the front panel knob and left/right arrow keys to enter the value (500 MHz)
as shown.
NOTE
If your microwave source does not have an ‘output enable/disable’ feature,
temporarily disconnect the cable from the microwave input channel until after
the zero-calibration is performed.