1
Performance Tests
44
Keysight 53210A/53220A/53230A Assembly Level Service Guide
2
With the external reference connected, use the following key sequence to
select (as necessary) the time base source to ‘Auto’ or ‘External’, and the
external reference frequency of 10 MHz:
3
Press the
Preset
key to preset the counter.
– Display shows:
–
Instrument has been preset.
–
Ext Ref
is shown in the upper right-hand corner of the display.
– The following softkeys are displayed:
4
Press and set:
– Measurement mode:
Ad vanced / Auto
*
*Opt 150
NOTE
The 53210A does not have an Auto measurement mode.