1
Performance Tests
76
Keysight 53210A/53220A/53230A Assembly Level Service Guide
8
Verify the ‘Mean’ is:
– 3.1 GHz ± 31 Hz (53210A)
– 3.1 GHz ± 3.1 Hz (53220A/53230A)
9
Verify the ‘Std Dev’ is less than:
– 43.4 Hz (53210A)
– 21.7 Hz (53220A)
– 4.34 Hz (53230A)
10
Record the Mean and Standard Deviation for 3.1 GHz, Test 6b,
in the Keysight 53210A/53220A/53230A Performance Test Record.
11
Press the
Reset Stats
softkey.
12
Set the E8257D-520 to 15.0 GHz at -4 dBm.
13
Press the
Trigger
key.
– The counter takes 100 readings and fills in the math functions.
14
Verify the ‘Mean’ is:
– 15.0 GHz ± 150 Hz (53210A)
– 15.0 GHz ± 15 Hz (53220A/53230A)
15
Verify the ‘Std Dev’ is less than:
– 210 Hz (53210A)
– 105 Hz (53220A)
– 21.0 Hz (53230A)
16
Record the Mean and Standard Deviation for 15.0 GHz, Test 6b,
in the Keysight 53210A/53220A/53230A Performance Test Record.
17
Press the
Reset Stats
softkey.
18
If Opt 150 is installed, proceed to Test 7c. Otherwise, disconnect the test
setup.
NOTE
The input power MUST be -4 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 15 GHz or a power meter to
verify input power.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...