Service
2
Keysight 53210A/53220A/53230A Assembly Level Service Guide
145
20
If any calibration step FAILS, after several seconds the counter displays
“Calibration step failed”
.
21
If the calibration fails (“
Calibration step failed
”), first verify the input frequency
and power are correct and the source output is connected and turned on.
Then, refer to
22
To check for error messages, refer to the
below for information on how to read the error
message(s) from a remote computer.
Table 2-3
Option 115 calibration frequencies and power levels
Test point
Frequency (MHz)
Power level 1
Power level 2
Power level 3
1
300 MHz
-21 dBm
-4 dBm
+10 dBm
2
1000 MHz
-21 dBm
-4 dBm
+10 dBm
3
3000 MHz
-21 dBm
-4 dBm
+10 dBm
4
5000 MHz
-21 dBm
-4 dBm
+10 dBm
5
7000 MHz
-21 dBm
-4 dBm
+10 dBm
6
9000 MHz
-21 dBm
-4 dBm
+10 dBm
7
10000 MHz
-21 dBm
-4 dBm
+10 dBm
8
11000 MHz
-21 dBm
-4 dBm
+10 dBm
9
12000 MHz
-21 dBm
-4 dBm
+10 dBm
10
13000 MHz
-21 dBm
-4 dBm
+10 dBm
11
14000 MHz
-21 dBm
-4 dBm
+10 dBm
12
15000 MHz
-21 dBm
-4 dBm
+10 dBm
NOTE
If the calibration fails for any calibration point (including zero-cal), the current
calibration constants for that frequency and power level, and the
Cal Count,
remain unchanged.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...