2
Service
116
Keysight 53210A/53220A/53230A Assembly Level Service Guide
To unlock the counter for calibration:
1
Press the
Utility
key, and then press the
Cal/AutoCal
softkey.
Unlock for calibration:
1
If
Unlock Cal
is displayed, use the counter front panel knob and the appropriate
left/right arrow keys to enter a valid security code into the data entry field on
and
– For example, if AT53210A was the security code set at the factory
(for a 53210A Counter), then enter
AT53210A
into the data entry field. The code
will repeat in the
Secure Code
line in the Calibrate Information screen.
2
Press the
Unlock Cal
softkey to complete the data entry.
– If the code is valid, the
Unlock Cal
annunciator changes to
Lock Cal
to show the
data was accepted and the counter is now unlocked for calibration.
– The
Cal Coun
t
is not incremented.
– If “AT53210A” is NOT the correct security code, the counter will display:
– “Incorrect Security Code - Try Again”
NOTE
If
Lock Cal
is displayed, then the counter is already unlocked for calibration, and
the remaining steps are not applicable.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...