Service
112
Keysight 53210A/53220A/53230A Assembly Level Service Guide
Figure 2-3
Calibration procedure summary
Calibration
Succeeded?
Press Perform
Cal Step
softkey.
Enter value of
source signal
using arrow keys
and knob.
Correct source
to be calibrated
shown on
Cal
Step
softkey?
Press
Perform
Cal
softkey.
Is Locked
softkey
visible?
Press
Cal/Autocal
softkey.
Press
Utility
key.
Connect Cal
Source to Channel
1 of counter.
Enter secure
code using
keys and knob.
Press
Unlock
softkey.
See Service
Manual.
Press
Cal Step
softkey.
Press a softkey:
Oscil
,
PosVolt
,
or
NegVolt
No
Yes
Yes
No
Yes
No
DONE: Press
any key to
continue
See Service
Manual
DONE: Press
any key to
continue
Calibration
Succeeded ?
No
Yes