Keysight 53210A/53220A/53230A Assembly Level Service Guide
121
Calibration cycle
The calibration cycle for the Keysight 53210A, 53220A, and 53230A Counters is
1-year. The counter should be calibrated at least once every 12 months to ensure
specification accuracy.
Calibration string:
The
Cal String
on the
Calibrate Information
page is operator-modifiable and can be
used to remotely enter any current or historical data/information up to 40
characters long. The string may include the characters A-Z, 0-9, and special
characters like @, #, *, etc.
The
CAL STRing
can be entered remotely as follows:
– Send CALibration:STRing “<string>” .
– (Example:) output <address> ; CAL:STR “last cal 8-15-2010”
– CALibration:STRing? reads the current cal string contents from the counter:
– “last cal 8-15-2010”
Internal reference oscillator calibration
1
First, run the counter Self-Test as described in
on page 169. A successful self-test procedure verifies that the
counter is operational and ready for calibration.
2
Unlock the counter by performing the procedure
[Utility] / (Cal AutoCal) / enter <code> / (Unlock Cal)
3
Set up a low-jitter, low-noise, external frequency source, such as the Stanford
Research CG635, for a 10 MHz square wave at 1 Vpp, 0 volts offset, 50% duty
cycle.
NOTE
If your source has an output enable feature, turn it ON.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...