2
Service
170
Keysight 53210A/53220A/53230A Assembly Level Service Guide
Summary of remote self-test verification
– Send command
*TST?
– Wait 10 seconds.
– Retrieve query result:
– 0 indicates the self-test PASSED.
– Any value other than 0 indicates self-test FAILED.
– If the retrieved value is non-zero, read error queue: SYSTem:ERRor?
– Read errors until no more errors.
To determine the counter firmware version
Older firmware versions may have known or unknown problems in the code. Verify
the firmware for your counter as follows:
1
Press the
Help
key, then the
About
softkey.
NOTE
If you are able to read the error message(s), consult
for what to do after receiving a specific calibration error message.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...