Performance Tests
1
Keysight 53210A/53220A/53230A Assembly Level Service Guide
87
14
Press to set:
– Statistics:
On
– The display shows:
Channel 3 pulse burst frequency test procedure
1
Press the
Trigger
key.
– The counter takes 100 readings and fills in the math functions.
2
Verify the ‘Mean’ is:
– 1.35 GHz ± 1.35 MHz
3
Verify the ‘Std Dev’ is less than:
– 338 kHz
4
Record the Mean and Standard Deviation for Test 7c, in the
53220A/53230A Performance Test Record (Tests 1 to 7)”
5
Press the
Reset Stats
softkey.