Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-102
2450-901-01 Rev. B/September 2013
:SYSTem:EVENtlog:POST
This command allows you to post messages to the event log.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:EVENtlog:POST <message>
:SYSTem:EVENtlog:POST <message>, <severityType>
<message>
A string that contains the message that will be associated with this event
<severityType>
The type of event that is generated; if nothing is defined, defaults to
INFormational
:
•
INFormational
•
ERRor
•
WARNing
Details
You can use this command to create your own event log entries and assign a severity level to them. This can be
useful for debugging and status reporting.
You must set the Log Warnings and Log Information options to be reported using the front panel to have the
custom warning and information events placed into the event log.
Example
SYST:EVEN:POST "my error", INF
SYST:EVEN:NEXT?
Posts an error named "my error."
Output:
1003,"User: my
error;4,1400469179,431599191"
Also see
None