Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-249
upgrade.unit()
This function upgrades the Model 2450 firmware.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
upgrade.unit()
Details
When
upgrade.unit()
is used, the firmware is only loaded if the version of the firmware component is newer
than the existing version. If the version is older or at the same revision level, it is not upgraded.
When you send this function, the instrument searches the flash drive that is inserted in the front-panel USB port
for an upgrade file. If the file is found, the instrument verifies that the file is a newer version. If the version is
older or at the same revision level, it is not upgraded. If it is a newer version, the instrument performs the
upgrade. An error is returned if no upgrade file is found.
Also see
(on page A-3)
(on page 8-248)
userstring.add()
This function adds a user-defined string to nonvolatile memory.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
userstring.add(
name
,
value
)
name
The name of the string; the key of the key-value pair
value
The string to associate with
name
; the value of the key-value pair
Details
This function associates the string
value
with the string
name
and stores this key-value pair in nonvolatile
memory.
Use the
userstring.get()
function to retrieve the
value
associated with the specified
name
.
You can use the
userstring
functions to store custom, instrument-specific information in the instrument, such
as department number, asset number, or manufacturing plant location.