
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2450-901-01 Rev. B/September 2013
3-45
Storing configuration point 2
Use one of the following methods to:
•
Set the instrument source level to 3.0 V
•
Store all active source settings to
MySourceList
as configuration point 2 by appending to the
end of the list
Using SCPI commands:
:SOURce:VOLTage:LEVel 3
:SOURce:CONF:LIST:STORe "MySourceList"
Using TSP commands:
smu.source.level = 3
smu.source.configlist.store("MySourceList")
The following figure shows the active state of the instrument after you change the source level to
Instrument settings stored in a source configuration list
(on page 3-38) for a complete
list of source settings that the instrument stores in a source configuration list.
Figure 83: Example configuration point 2