The
trigger.blender[
N
].reset()
function resets the following attributes to their factory defaults:
•
trigger.blender[
N
].orenable
•
trigger.blender[
N
].stimulus[
M
]
It also clears
trigger.blender[
N
].overrun
.
Example
trigger.blender[1].reset()
Resets the trigger blender 1
settings to factory defaults.
Also see
(on page 8-166)
(on page 8-167)
trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script trigger.EVENT_NONE
Usage
event
= trigger.blender[
N
].stimulus[
M
]
trigger.blender[
N
].stimulus[
M
] =
event
event
The event that triggers the blender action; see
Details
N
An integer that represents the trigger event blender (1 or 2)
M
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event. The
event
parameter can be any trigger
event.