Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 2: General operation
2450-901-01 Rev. B/September 2013
2-109
Setting autozero
You can enable or disable automatic referencing, or request a one-time refresh of the reference
values.
The reference setting is stored with the measure function.
To set autozero using the front panel:
1. Press the
FUNCTION
key.
2. Select the source and measurement combination.
3. Press the
MENU
key.
4. Under Measure, select Settings.
5. Next to Auto Zero, select
On
or
Off
.
6. If Off is selected, you can select the
Once Now
option to send a one-time refresh. When you
select Once Now, Auto Zero is enabled after the refresh.
7. Select
HOME
to return to the operating display.
To set autozero using SCPI commands:
To turn autozero on, send the command:
:SENSe:VOLTage:AZERo ON
To turn autozero off, send the command:
:SENSe:VOLTage:AZERo OFF
To set autozero on or off for current measurements, replace
VOLTage
with
CURRent
. To set it for
resistance measurements, replace it with
RESistance
.
To perform a one-time autozero update, send the command:
:SENSe:AZERo:ONCE
To set autozero using TSP commands:
To turn autozero on, select the measurement type, then send the command:
smu.measure.autozero.enable = smu.ON
To turn autozero off, select the measurement type, then send the command:
smu.measure.autozero.enable = smu.OFF
To perform a one-time autozero update, select the measurement type, and then send the command:
smu.measure.autozero.once()