Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-154
2450-901-01 Rev. B/September 2013
:TRIGger:LAN<n>:IN:CLEar
This command clears the event detector for a trigger.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
TRIGger:LAN<n>:IN:CLEar
<n>
The LAN event number (1 to 8)
Details
The trigger event detector enters the detected state when an event is detected. This function clears a trigger
event detector and discards the previous history of the trigger packet.
This function clears all overruns associated with this LAN trigger.
Example
TRIG:LAN5:IN:CLE
Clears the event detector with LAN packet 5.
Also see
None