Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-134
2450-901-01 Rev. B/September 2013
Example
CALC2:LIM1:STAT ON
CALC2:LIM1:LOW -5.17
CALC2:LIM1:UPP -4.23
TRIG:BLOC:BRAN:LIM:DYN 9, IN, 1, 12, 7
Set the limits on with a low limit of -5.17 and
a high limit of -4.23. Set trigger block 9 to
test if the limit is inside those limits based on
the measurement reading at block 7. If the
measurement is within the limits, go to block
12.
This example is a snippet of a larger
example. For the complete code, see
grading with immediate binning example
Also see
(on page 3-100)
:CALCulate2:<function>:LIMit<y>:LOWer[:DATA]
(on page 6-16)
:CALCulate2:<function>:LIMit<y>:UPPer[:DATA]
(on page 6-18)
:TRIGger:BLOCk:BRANch:ONCE
This command causes the trigger model to branch to a specified building block the first time it is encountered in
the trigger model.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:ONCE <blockNumber>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<branchToBlock>
The block number of the trigger model block to execute when the trigger
model first encounters this block
Details
The branch-once building block branches to a specified block the first time the trigger model encounters the
branch-once block. If it is encountered again, the trigger model ignores the block and continues in the normal
sequence.
The once block is reset when the trigger model reaches the idle state. Therefore, the branch-once block always
executes the first time the trigger model encounters this block.
Example
:TRIG:BLOC:BRAN:ONCE 2, 4
When the trigger model reaches block 2, the trigger model goes to block 4 instead of proceeding to the default
sequence of block 3.
Also see
(on page 3-100)