Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-34
2450-901-01 Rev. B/September 2013
:OUTPut[1][:STATe]
This command enables or disables the source output.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
0 (OFF)
Usage
:OUTPut[1][:STATe] <b>
:OUTPut[1][:STATe]?
<b>
Turn source off:
0
or
OFF
Turn source on:
1
or
ON
Details
When the output is switched on, the instrument sources either voltage or current, as set by
[:SOURce[1]]:FUNCtion[:MODE]
.
Example
:OUTP ON
Switch the source output of the instrument to
on.
Also see
(on page 6-72)