Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-43
[:SENSe[1]]:<function>:AVERage:COUNt
This command sets the number of measurements that are averaged when filtering is enabled for the function
measurements.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
10
Usage
[:SENSe[1]]:<function>:AVERage:COUNt <n>
[:SENSe[1]]:<function>:AVERage:COUNt?
[:SENSe[1]]:<function>:AVERage:COUNt? DEFault
[:SENSe[1]]:<function>:AVERage:COUNt? MINimum
[:SENSe[1]]:<function>:AVERage:COUNt? MAXimum
<function>
The measurement function to which this setting applies:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<n>
The number of readings required for each filtered measurement (1 to 100)
Details
The filter count is the number of readings that are acquired and stored in the filter stack for the averaging
calculation. The larger the filter count, the more filtering that is performed.
If you send this command without the
<function>
parameter, it sets the filter count for all functions.
Example 1
CURR:AVER:COUNT 10
CURR:AVER:TCON MOV
CURR:AVER ON
For current measurements, set the averaging
filter type to moving average, with a filter count of
10.
Enable the averaging filter.
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
For resistance measurements, set the averaging
filter type to moving average, with a filter count of
10.
Enable the averaging filter.
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
For voltage measurements, set the averaging
filter type to moving average, with a filter count of
10.
Enable the averaging filter.
Also see
(on page 4-22)
[:SENSe[1]]:<function>:AVERage[:STATe]
(on page 6-44)
[:SENSe[1]]:<function>:AVERage:TCONtrol
(on page 6-45)