Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-111
smu.measure.math.enable
This command enables or disables math operation on measurements for the selected measurement function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.OFF
Usage
value
= smu.measure.math.enable
smu.measure.math.enable =
value
value
The math enable setting:
•
Enable:
smu.ON
or
1
•
Disable:
smu.OFF
or
0
Details
When this command is set to on, the math operation specified by the math format command is performed before
completing a measurement.
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.math.format =
smu.MATH_PERCENT
smu.measure.math.enable = smu.ON
When voltage measurements are made, the math
format is enabled and set to percent.
Also see
Calculations that you can apply to measurements
(on page 3-6)