Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-42
2450-901-01 Rev. B/September 2013
[:SENSe[1]]:COUNt
This command sets the number of measurements to be performed when a measurement is requested.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
[:SENSe[1]]:COUNt <n>
[:SENSe[1]]:COUNt DEFault
[:SENSe[1]]:COUNt MINimum
[:SENSe[1]]:COUNt MAXimum
[:SENSe[1]]:COUNt?
[:SENSe[1]]:COUNt? DEFault
[:SENSe[1]]:COUNt? MINimum
[:SENSe[1]]:COUNt? MAXimum
<n>
The number of measurements (1 to 300,000)
Details
This command sets the number of measurements that are taken when a measurement is requested. This
command does not affect the trigger model.
To get better feedback from the instrument, use the Simple Loop trigger model template instead of
using the count command.
Example
COUN 10
Make ten measurements.
Also see
(on page 6-4)