Section 3: Functions and features
Model 2450 Interactive SourceMeter® Instrument Reference Manual
3-50
2450-901-01 Rev. B/September 2013
Using SCPI commands:
To receive the name of one source configuration list stored on the instrument, use the following
command.
:SOURce:CONFigure:LIST:CATalog?
For a measure configuration list, replace
:SOURce
with
:SENSe
.
Each time this command executes, the name of one defined configuration is returned. Keep sending
this command until it returns an empty string to get all defined lists. After the command returns an
empty string, it wraps around and starts returning names again. If only an empty string is returned, no
configuration lists of the specified type exist.
Using TSP commands:
To receive the name of one source configuration list stored on the instrument, use the following
command.
print(smu.source.configlist.catalog())
For a measure configuration list, replace
source
with
measure
.
Each time this command executes, the name of one defined configuration is returned. Keep sending
this command until it returns
nil
to get all defined lists. After the command returns
nil
, it wraps
around and starts returning names again. If only
nil
is returned, no configuration lists of the specified
type exist.
Determining the size of a configuration list
You can view the number of configuration points that are in a specific configuration list.
Using SCPI commands:
To view the number of configuration points in a source configuration list named
MyConfigList
,
send the following command:
:SOURce:CONFiguration:LIST:SIZE? "MyConfigList"
For a measure configuration list, replace
:SOURce
with
:SENSe
.
Using TSP commands:
To view the number of configuration points in a source configuration list named
MyConfigList
,
send the following command:
smu.source.configlist.size("MyConfigList")
For a measure configuration list, replace
source
with
measure
.
Saving a configuration list
Configuration lists are lost when you turn the instrument off and turn it on again. Save a configuration
list by creating a configuration script (TSP or front panel) or using the
*SAV
and
*RCL
commands
(SCPI). A configuration script saves the settings of the instrument, including all defined source and
measure configuration lists. See
(on page 2-112) for additional information.