Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2450-901-01 Rev. B/September 2013
3-59
Logarithmic sweep with a current source
The following examples perform a logarithmic sweep using a current source. They perform the
following actions:
•
Reset the instrument to its defaults.
•
Set the source function to current.
•
Set the source range to 100 mA.
•
Set up a logarithmic sweep from 100
µ
A to 100 mA in 10 steps with a source delay of 10 ms, a
sweep count of 1, and a fixed source range. In TSP only, name the configuration list that is
created for this sweep
CurrLogSweep
.
•
Set the measure function to current.
•
Set the current range to 100
µ
A.
•
Start the sweep.
No buffer is defined, so the data is stored in
defbuffer1
(on page 3-10) for
more information on reading buffers.
Using SCPI commands
*RST
SOUR:FUNC CURR
SOUR:CURR:RANG 100e-3
SENS:FUNC "VOLT"
SENS:VOLT:RANG 20
SOUR:SWE:VOLT:LOG 100e-6, 100e-3, 10, 10e-3, 1, FIXED
INIT
Using TSP commands
reset()
smu.source.func = smu.FUNC_DC_CURRENT
smu.source.range = 100e-3
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.range = 20
smu.source.sweeplog("CurrLogSweep", 100e-6, 100e-3, 10, 10e-3, 1, smu.RANGE_FIXED)
trigger.model.initiate()