Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2450-901-01 Rev. B/September 2013
3-3
To set the relative offset for another function, replace
VOLTage
with
CURRent
or
RESistance
.
Using TSP commands:
Send the commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.rel.level =
relValue
smu.measure.rel.enable = smu.ON
Where
relValue
is the relative offset value.
To set the relative offset for another function, replace
smu.FUNC_DC_VOLTAGE
with
smu.FUNC_DC_CURRENT
or
smu.FUNC_RESISTANCE
.
Disabling the relative offset
Using the front panel:
1. Select the measurement function to which the relative offset is applied.
2. On the SETTINGS swipe screen, select the box next to
Rel
. An X should be displayed and the
REL annunciator to the right of the measurement is no longer displayed.
The relative offset feature can also be disabled by selecting MENU > Filter/Math > Rel State and
selecting Off.
Using SCPI commands:
Send the command:
:SENSe:VOLTage:RELative OFF
To set the relative offset for another function, replace
VOLTage
with
CURRent
or
RESistance
.
Using TSP commands:
Send the commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.rel.enable = smu.OFF
To set the relative offset for another function, replace
smu.FUNC_DC_VOLTAGE
with
smu.FUNC_DC_CURRENT
or
smu.FUNC_RESISTANCE
.