Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-162
2450-901-01 Rev. B/September 2013
:TRIGger:LOAD:EMPTy
This command resets the trigger model.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:LOAD:EMPTy
Details
When you load this predefined trigger model, any existing trigger model settings are reset.
Any existing trigger blocks are deleted when you execute this command.
Example
TRIG:LOAD:EMPTY
TRIG:BLOC:BUFF:CLEAR 1
TRIG:BLOC:MEAS 2
TRIG:BLOC:BRAN:COUN 3, 5, 2
TRIG:BLOC:DEL:CONS 4, 1
TRIG:BLOC:BRAN:COUN 5, 3, 2
Reset trigger model settings.
Clear
defbuffer1
at the beginning of execution
of the trigger model.
Loop and take 5 readings.
Delay a second.
Loop three more times back to block 2.
At the end of execution, 15 readings are stored
in
defbuffer1
.
Also see
None