Section 8: TSP command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
8-104
2450-901-01 Rev. B/September 2013
smu.measure.limit[Y].autoclear
This command indicates if limit
Y
should be cleared automatically or not.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.ON
Usage
value
= smu.measure.limit[
Y
].autoclear
smu.measure.limit[
Y
].autoclear =
value
value
The auto clear setting:
•
Enable:
smu.ON
or
1
•
Disable:
smu.OFF
or
0
Y
Limit number:
1
or
2
Details
When this command sets autoclear to on for a measurement function, if a measurement fails limit, but the next
measurement passes limit, the failed limit condition is cleared. Therefore, if you are making a series of
measurements, the instrument uses last measurement limit for the fail indication for the limit.
If you want to know if any of a series of measurements failed the limit, set the auto clear setting to off. When this
set to off, a failed indication is not cleared automatically. It remains set until it is cleared with the clear command.
The auto clear setting affects both the high and low limits of
Y
.
Example
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.limit[1].autoclear = smu.ON
Turns on autoclear for limit 1 when
measuring DC current.
Also see
(on page 8-105)