Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-37
[:SENSe[1]]:CONFiguration:LIST:CATalog?
This command returns the name of one measure configuration list.
Type
Affected by
Where saved
Default value
Query
Not applicable
Not applicable
Not applicable
Query
[:SENSe[1]]:CONFiguration:LIST:CATalog?
Details
You can use this command to see the names of measure configuration lists stored on the instrument.
This command returns one name each time you send it. This command returns an empty string when there are
no more names to return. If the command returns an empty string the first time you send it, no measure
configuration lists have been created for the instrument.
Examples
:SENSe:CONFiguration:LIST:CATalog?
Send this command to receive the name of one measure
configuration list stored on the instrument. Keep sending this
command until it returns
NULL
to get all stored lists.
Also see
(on page 3-33)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-37)
[:SENSe[1]]:CONFiguration:LIST:CREate
This command creates an empty measure configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:CREate <name>
<name>
A string that represents the name of a measure configuration list
Details
This command creates an empty configuration list. To add configuration points to this list, you need to use the
store command.
Configuration lists are not saved when the instrument is turned off. If you want to save a configuration list, use a
saved setup to store the instrument settings, which include defined configuration lists.
Example
:SENS:CONF:LIST:CRE "MyMeasList"
Creates a measure configuration list named
MyMeasList
.
Also see
(on page 6-2)
(on page 3-33)
[:SENSe[1]]:CONFiguration:LIST:STORe
(on page 6-41)