Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-117
Details
The filename must specify the full path (including
/usb1/
). If included, the file extension must be set to
.csv
(if
no file extension is specified,
.csv
is added).
For options that save more than one item of time information, each item is comma-delimited. For example, the
default format is
date
,
time
, and
fractional seconds
for each reading.
Example
TRACe:MAKE "MyBuffer", 100
SENSe:COUNt 5
MEASure:CURRent:DC? "MyBuffer"
TRACe:DATA? 1,5, "MyBuffer", READ, REL, SOUR
TRACe:SAVE "/usb1/myData.csv", "MyBuffer"
TRACe:SAVE "/usb1/myDataRel.csv", "MyBuffer", REL
Create a buffer called
MyBuffer
with a
maximum size of 100.
Make five readings for each
measurement request and return the
data.
Make the measurements.
Read the readings, relative timestamp,
and source value for each point from 1
to 5.
Output:
-0.000000,0.000000,
0.000000,-0.000000,
0.301759,0.000000,
-0.000000,0.579068,
0.000000,-0.000000,
0.884302,0.000000,
-0.000000,1.157444,
0.000000
Save all reading and default time
information from a buffer named
MyBuffer
to a file named
myData.csv
on the USB flash drive.
Save all readings and relative time
stamps from
MyBuffer
to a file named
myDataRel.csv
on the USB flash drive.
Also see
(on page 3-10)
(on page 3-27)
(on page 6-114)
(on page 6-118)