Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-135
:TRIGger:BLOCk:BRANch:ONCE:EXCLuded
This command causes the trigger model to go to a specified building block every time the trigger model
encounters it, except for the first time.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:ONCE:EXCLuded <blockNumber>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<branchToBlock>
The block number of the trigger model block to execute when the trigger
model encounters this block after the first encounter
Details
The branch-once-excluded building block is ignored the first time the trigger model encounters it. After the first
encounter, the trigger model goes to the specified branching block.
The branch-once-excluded block is reset when the trigger model starts.
Example
:TRIG:BLOC:BRAN:ONCE:EXCL 2, 4
When the trigger model reaches block 2 the first time, the trigger model goes to block 3. If the trigger model
reaches this block again, the trigger model goes to block 4.