Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2450-901-01 Rev. B/September 2013
3-101
Once excluded
The branch-once-excluded building block is ignored by the trigger model the first time it is
encountered. If the trigger model encounters the block again, it goes to a specified building block.
The branch-once-excluded block is reset when the trigger model starts.
Trigger block summary
Front-panel icon
SCPI command
TSP command
Block description
Not applicable
(on page 6-142)
(on page 8-186)
This returns the
settings for all trigger
model building
blocks
(on page 6-128)
trigger.model.setblock() — trigger.BLOCK_BRANCH_ALWAYS
This defines a
trigger model block
that always goes to
a specific block
Not applicable
(on page 6-128) (query)
trigger.model.getbranchcount()
(on page 8-187)
This returns the
count value of the
trigger model
counter block
(on page 6-128)
trigger.model.setblock() — trigger.BLOCK_BRANCH_COUNTER
This branches to a
specified block a
specified number of
times
(on page 6-130)
trigger.model.setblock() — trigger.BLOCK_BRANCH_DELTA
(on page
This defines a
trigger model block
that goes to a
specified block if the
difference of two
measurements
meets preset criteria