Model 2450 Interactive SourceMeter® Instrument Reference Manual
Appendix D: Model 2450 in a Model 2400 application
2450-901-01 Rev. B/September 2013
D-17
Model 2400 command
:SOURce[1]:CURRent:CENTer
:SOURce[1]:CURRent:CENTer?
:SOURce[1]:VOLTage:CENTer
:SOURce[1]:VOLTage:CENTer?
Model 2450 command
:SOURce[1]:SWEep:<function>:LINear
(on page 6-81)
:SOURce[1]:SWEep:<function>:LINear:STEP
(on page 6-83)
:SOURce[1]:SWEep:<function>:LIST
(on page 6-85)
:SOURce[1]:SWEep:<function>:LOG
(on page 6-87)
Notes
Sweep parameters are built into the sweep command path. See
(on page 3-51) for information.
Model 2400 command
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude]
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude]?
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Model 2450 command
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude]
(on page 6-70), where
<function>
is
CURRent
or
VOLTage
.
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Notes
Model 2400 command
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude]
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude]?
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Model 2450 command
Not available
Notes
Use the
(on page 3-95).
Model 2400 command
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor?
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor?
Model 2450 command
Not available
Notes
Use the
(on page 3-95).
Model 2400 command
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor:STATe
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor:STATe?
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor:STATe
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor:STATe?
Model 2450 command
Not available
Notes
Use the
(on page 3-95).
Model 2400 command
:SOURce[1]:CURRent:MODE
:SOURce[1]:CURRent:MODE?
:SOURce[1]:VOLTage:MODE
:SOURce[1]:VOLTage:MODE?
Model 2450 command
:SOURce[1]:FUNCtion[:MODE]?
Notes