Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-41
Details
This command returns the size (number of configuration points) of a measure configuration list..
The size of the list is equal to the number of configuration points in a configuration list.
Examples
:SENSe:CONF:LIST:SIZE? "MyMeasList"
Returns the number of configuration points in
a measure configuration list named
MyMeasList
.
Also see
(on page 3-33)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-37)
[:SENSe[1]]:CONFiguration:LIST:STORe
This command stores the active measure settings into the named configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Saved settings
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:STORe <name>
[:SENSe[1]]:CONFiguration:LIST:STORe <name>, <point>
<name>
A string that represents the name of a measure configuration list
<point>
A specific configuration point in the configuration list
Details
Use this command to store the active settings to a configuration point in a configuration list. If you do not include
the <point> parameter, the configuration point will append to the end of the list.
Refer to
Instrument settings stored in a measure configuration list
(on page 3-36) for a complete list of measure
settings that the instrument stores in a measure configuration list.
Example
:SENSe:CONF:LIST:STOR "MyConfigList"
Stores the active settings of the instrument to the
end of the configuration list named
MyConfigList
.
:SENSe:CONF:LIST:STOR "MyConfigList", 5
Stores the active settings of the instrument to the
configuration list named
MyConfigList
in
configuration point 5.
Also see
(on page 3-33)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-37)