DIGISPENSE 3009 CONTROLLER MODULE, STYLE B
Page 3-92
29 October, 2019
IVEK Corp. P03R017E
c => CLEAR FAULTS
c or c0
c0,<value2>
Clears all faults and reference pumps.
d => DIRECTION
d or d0
d0,<value2>
Returns current Production Mode fluid direction setting.
d0,<value2>
d0,<value2>
Sets the Production Mode fluid direction.
<value2>:
0 = Reverse
1 = Forward (default)
d1
d1,<value2>
Returns current Fluidic Mode direction setting.
d1,<value2>
d1,<value2>
Sets the Fluidic Mode direction.
<value2>:
0 = Reverse
1 = Forward (default)
e => END
e or e0
e0
Ends the current pumping cycle.
In Prime mode, will continue until piston has reached the stop position.
f => REFERENCE
f or f0
f0
References the pump.
g => TOTALIZER
g or g0
g0,<value2>
Returns the totalizer value in number of Strokes.
g0,<value2>
g0,<value2>
Resets the totalizer
value to 0.
<value2>: 0 = Resets the totalizer value to zero.
g1
g1,<value2>
Returns the totalizer value in number of Cycles.
g1,<value2>
g1,<value2>
Resets the totalizer
value to 0.
<value2>: 0 = Resets the totalizer value to zero.
g2
g2,<value2>
Returns the last prime volume/agitate volume read in strokes.
MAX VALUE:
The totalizer will increment to a maximum value of 3,999,999,999 then 'wrap around’.
h => HARDWIRED READY SIGNAL OPERATION
h1
h1,<value2>
Returns the Controller address for serial interface setting.
h1,<value2>
h1,<value2>
Sets the Controller address for serial interface configuration.
<value2>
1 = Minimum (default)
99 = Maximum
h2,<value2>
h2,<value2>
Sets the contrast percentage.
<value2>
30 = Minimum
50 = (default)
70 = Maximum
h3,<value2>
h3,<value2>
Enables/Disables Comm Gateway mode.
<value2>
0 = Disabled (default)
1 = Enabled)
h4,<value2>
h4,<value2>
Enabled/Disabled Logic I/O Gateway mode.
<value2>
0 = Disabled (default)
1 = Enabled
h5,<value2>
h5,<value2>
Configures the conduction level for Recipe Outputs when 'true'..
<value2>
0 = Recipe Outputs conducting when 'true'.
1 = Recipe Outputs non-conducting when 'true'.
h10
h10,<value2> Returns the Input CC 1 setting.
h10,<value2> h10,<value2> Sets the Input CC 1 configuration.