
DIGISPENSE 3009 CONTROLLER MODULE, STYLE B
Page 3-46
29 October, 2019
IVEK Corp. P03R017E
Dispense Screen Description
- The Dispense screen displays the current Dispense parameters and allows a user
with
Supervisor permission to change the parameters. The current permission level determines which push-buttons
are available as shown.
Status -
The status indicates the operational state of the system. Refer to section 3.3.8 for an explanation of each
status
type.
Total Volume -
Total Volume is the total number of strokes of the motor in Dispense (and Meter) modes. The total
strokes
counter will go up to 3,999,999,999 before restarting at 0.
Total Cycles -
Total Cycles is the total number of completed dispense/meter cycles in Dispense (and Meter) modes.
The
total cycles counter will go up to a maximum of 3,999,999,999 before restarting at 0.
Recipe -
The Recipe line shows the current recipe selected in the Recipe screen. "0" indicates values changed since
last
"Get Recipe".
Direction -
The Direction is the direction of liquid flow through the Pump Module for Dispense (and Meter) modes.
The direction
is either Forward or Reverse.
Dispense Volume -
The Dispense Volume is the number of strokes required for a dispense. The range is 1 to 10,000
with
an increment value of 1.
Dispense/Meter Rate -
The Dispense/Meter Rate is the speed of the motor in revolutions per minute (RPM) during
the
Dispense (and Meter) operation. The range is 1 RPM to 1,500 RPM with an increment value of 1 and is
dependent on the Motor/Base type.
Drawback -
Selecting Drawback will bring you to the Drawback screen. If the Drawback Volume setting is 0.00 the
Meter
screen will show drawback as disabled. Any other Volume setting will show drawback as Enabled.
To change a value;
1. Press the Arrow push-buttons to highlight the parameter.
2. Press the Enter push-button to bring up either the Select New Value or Enter New Value screen.
3. In the Select New Value screen use the up and down arrow push-buttons to select the value. In the Enter New
Value screen press the Minimum or Maximum push-buttons to select the desired value or use the left and right
arrow push-buttons to select the digit and the up and down arrow push-buttons to change the digit's value.
4. Press the Enter push-button.
NOTE
Changing Direction, Dispense/Meter Rate or Drawback does not affect the settings of Prime or Agitate mode.
However, they do affect Meter mode.