Ecotron - Incubation Shaker
Rectifying Faults
Page
66
of
90
03 May 2018
Fault
The shaker drive accelerates too quickly or runs too fast.
Possible cause
Measures
Personnel
Error when measuring the ro-
tation speed.
Switch the equipment off and then on again.
Operator
The drive control system is
defective.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Operator
Faults related to the “Temperature” parameter
Fault
The temperature fails to reach the desired setpoint.
Possible cause
Measures
Personnel
The heating is not switched
on.
Activate the “Temperature” parameter.
Operator
The setpoint is set too low.
Raise setpoint.
Operator
The ambient temperature
does not meet the require-
ments.
see chapter 6.1, page 41
Operator
Door not closed completely.
Close door completely.
Operator
Cooling fins dusty.
Clean the cooling fins.
Operator
Error during temperature
measurement.
Check the function of the temperature sensor. In
case of a defect, contact your local Infors repre-
sentative.
Operator
Heater fan not working.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Infors service techni-
cian or licensed dealer
INFORMATION
Re-measuring the temperature only provides reliable information
when using calibrated measuring tools, and only if they are used
at points specified by Infors. Measurements taken at undefined
points in the casing will not produce any usable data.
For information about temperature measuring, contact your local
Infors representative or request a quote for calibrating the pa-
rameters.