Ecotron - Incubation Shaker
Rectifying Faults
Page
64
of
90
03 May 2018
Alarm message
Meaning
The rotation speed is too high.
Possible cause
Measures
Personnel
Error when measuring the ro-
tation speed.
Switch the equipment off and then on again.
Operator
The drive control system is
defective.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Operator
The motor is defective.
Contact your local Infors representative.
Infors service techni-
cian or licensed dealer
Alarm message
Meaning
The equipment door is open.
8.2 Fault Tables
The tables below describe potential faults that do not usually trig-
ger an error message on the display or – with a few exceptions –
an alarm signal. In many cases, operators can resolve these faults
themselves.
General faults
Fault
After activating the main switch, the view boxes do not light up.
Possible cause
Measures
Personnel
Power supply to equipment in-
terrupted.
Check if the plugs are plugged in correctly.
Check the mains connection.
Operator
The equipment fuse is defec-
tive.
Replace equipment fuse (see chapter 10.1, page
72).
Operator
Fault
The interior lighting does not work.
Possible cause
Measures
Personnel
The lighting element of the in-
terior lighting is defective.
Replace the lighting element (see chapter 10.3,
page 74).
Operator