Ecotron - Incubation Shaker
Rectifying Faults
03 May 2018
Page
65
of
90
Faults related to the “Rotation speed” parameter
Fault
The table does not move.
Possible cause
Measures
Personnel
The “Rotation speed” parame-
ter is not activated.
Activate the parameter.
Operator
The table is blocked by for-
eign objects.
Dismantle table and remove foreign objects.
Operator
The load is too heavy, making
the drive too weak or causing
it to overheat.
Reduce the load.
If the drive has overheated, the process can be
continued after it has cooled down.
Operator
The drive has heated up too
much, and has been turned
off automatically.
Let the equipment cool down. Ensure that the air
vents on the side are unobstructed.
Operator
The cable from the shaker to
the control rack has been un-
plugged.
Plug in the cable.
Operator
The drive control system is
defective.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Operator
The drive belt is torn.
Contact your local Infors representative and have
the drive belt replaced.
Infors service techni-
cian or licensed dealer
The motor is defective.
Contact your local Infors representative.
Infors service techni-
cian or licensed dealer
Fault
Strong vibrations occur.
Possible cause
Measures
Personnel
The load is too heavy, so the
equipment is not balanced.
Reduce the load to the prescribed value.
Operator
The load is too light, so the
equipment is not balanced.
Increase the load to the prescribed value.
Operator
The load is distributed une-
venly.
Load the tray in the centre. If possible, do not
place any heavy weights in the corners of the
tray.
Operator
The rotation speed is too high. Reduce the rotation speed.
Operator
The equipment is not standing
straight.
Place the equipment on an flat surface.
Operator
The substructure (table) is too
weak.
Place the equipment on a stable table or the
floor.
Operator
The liners of the table are
worn.
Contact your local Infors representative and re-
place the liners (see chapter 10.5, page 76).
Operator