Ecotron - Incubation Shaker
Operation
Page
58
of
90
03 May 2018
7.4.3 Activating and Deactivating Parameters
Activating parameters
To activate a parameter, proceed as follows:
Procedure
1.
Press the
On/Off
button (1) to activate the desired parameter
(rotation speed on the left, temperature on the right).
The parameter is activated immediately. The display briefly
shows the message
ON
and then the current actual value of
the parameter.
When the “Rotation speed” (RPM) parameter is activated, the
shaking procedure starts up immediately. As such, we recommend
activating this parameter last. When starting up the shaker drive,
use the time until the setpoint is reached to check whether the cul-
tivation flasks are secured on the tray.
INFORMATION
The parameters need a little time to reach their set setpoints.
Until they have done so, the notice
Hi
or
Lo
is displayed, de-
pending on whether the actual value is higher or lower than the
setpoint.
Deactivating parameters
The parameters are deactivated using the
On/Off
button in the
same way as for activation. When a parameter is deactivated, the
view box alternates between the set setpoint and the message
OFF
.