Ecotron - Incubation Shaker
Operation
03 May 2018
Page
47
of
90
7 Operation
7.1 Switching on the Equipment
ATTENTION
The equipment can start up automatically if this option has not
been deactivated correctly beforehand. Any objects flying
around the incubation chamber can damage the equipment and
the cultivation vessels if the equipment starts up automatically.
– Prior to switching on the equipment, ensure that there are
no foreign objects in the incubation chamber.
Procedure
1.
Activate the main switch on the front.
If the equipment has been correctly connected to the power,
the
Power supply on/off
function symbol lights up after switch-
ing on.
Once it has been switched on, the equipment will auto-
matically run a self-test and the installed firmware version is
displayed briefly. Following that, the equipment is ready for
operation.
7.2 Loading the Equipment
CAUTION
If a vessel breaks at high rotation speed, glass splinters are
slung around the incubation chamber. When the door is open,
the equipment does not stop immediately. Therefore pieces of
glass can be slung out.
To avoid injuries:
– Do not open the door in a panic if you notice that a vessel
has broken.
– Switch off the equipment at the main switch and only open
the door when the table stands still.